How to open your own X-ray room: recommendations, instructions. Project for an X-ray medical office and commissioning

SK Olimp LLC offers turnkey design services for X-ray medical rooms in Moscow. Our experts will do everything necessary work for the creation and commissioning of the X-ray room, including the preparation of a full package of documents and obtaining licenses, conclusions and permits.

Find out the cost of the service - send a request for a project

What is included in the design services of medical X-ray rooms?

The creation of an X-ray room begins with the choice of a room and an X-ray machine. At this stage, it is already necessary to know some of the features that will be presented by Rospotrebnadzor in the future when putting the X-ray room into operation and obtaining a sanitary and epidemiological conclusion (SEZ).

The next step is the design of an X-ray room with the obligatory calculation of X-ray protection. Our company develops a technological project for an X-ray room as part of comprehensive training x-ray room "turnkey", and as a separate independent service. More information about X-ray room design can be found

The design of an X-ray room begins with a survey of the structural elements of the building and internal existing engineering systems. Then a package of documents is created, which includes text and graphic parts as part of:

  1. Explanatory note.
  2. Technological part (with the provision of calculation of protection from sources of ionizing radiation and radiation safety). Perform the calculation of radiation protection in accordance with the current standards (SanPiN 2.6.1.1192-03).
  3. Architectural and construction solutions.
  4. Structural and space-planning solutions.
  5. Estimated documentation.
  6. Information about engineering equipment, about networks of engineering and technical support, a list of engineering and technical measures:
  • subsection "Power supply system";
  • subsection “Water supply and sanitation system”;
  • subsection "Heating, ventilation";
  • subsection "Communication networks";
  • subsection "Automation".

Based on the design of the X-ray room, all subsequent repair and construction work is carried out, including the installation of X-ray protective structures and elements. SK OLIMP LLC offers a full range of construction, repair and installation works, including the creation of X-ray protection on the basis of a technological project agreed in the RRO and a license for the placement of sources of ionizing radiation (IRS), design and manufacture of radiation protection means of RRS (License No. 77.99.15.002.L.000059.08.14).

What data and requirements are taken into account when designing X-ray rooms?

Design documentation for medical rooms with X-ray equipment is developed in accordance with the Government Decree No. 87 of February 16, 2008 and the letter of the Ministry of Regional Development of the Russian Federation of June 22, 2009 No. 19088-sk / 08; RMD 11-08-2009. The composition and area of ​​the premises are designed on the basis of the results of measurement work and surveys (including an examination of water supply and water curing systems, ventilation, heating power supply), taking into account the planned functional purpose of the premises and the parameters set by the equipment manufacturer, and the requirements of SanPin 2.1.3.2630-10, SNiP 31-06-2009, SNiP 21-01-97, SP 2.6.1.2612-10, SanPiN 2.6.1.1192-03, TSN 31-330-2005, SP 31-110-2003, SNiP, SanPiN, PUE. The project must take into account the requirements of the Guidelines for the design of healthcare institutions, SanPiN 2.1.3.2630-10, SNiP 41-01-2003. Decree of the Government of the Russian Federation of April 25, 2012 N 390 "On the fire regime" and other applicable regulatory documents.

The planning solution for the re-equipment of the premises is carried out in accordance with the norms of technical regulations, SaNPiN, taking into account the location of the premises for the installation of an X-ray apparatus.

Sections of the working documentation of architectural and planning and architectural and construction solutions are developed taking into account the state of the supporting structures of the building in accordance with current rules operation of the premises.

Procedure for commissioning X-ray rooms

  • 1 Our company carries out "on a turn-key basis" the whole complex of works on design and repair and construction work with subsequent commissioning and obtaining a free economic zone. In cases where there are activities that involve not only the use of medical equipment for personal purposes, we provide a service for accompanying and obtaining a Rospotrebnadzor License for sources of ionizing radiation. More information
    The procedure for commissioning X-ray rooms is regulated by SanPiN 2.6.1.1192-03, as well as SP 2.6.1.1283-03, SanPiN 2.1.3.2630-10. According to SanPiN 2.6.1.1192-03, the documented commissioning process is as follows:
  • 2

    Purchase of an X-ray machine and receipt of documents provided by the equipment manufacturer:

    • Sanitary and epidemiological conclusion for the X-ray machine;
    • Registration certificate of the Ministry of Health of Russia for an X-ray machine (certified copy);
    • Operational documentation for the X-ray machine.

    You should be aware that not all types of dental x-ray machines can be used in clinics located in residential buildings. It depends on the purpose of the device, its workload and on the nominal value of the anode voltage. Restrictions on these parameters are specified in SanPiN 2.6.1.1192-03 "Hygienic requirements for the design and operation of X-ray rooms, apparatus and X-ray examinations."

  • 3

    Obtaining a license of an institution for medical activities.

  • 4

    Please note that there are several options for placing the X-ray machine:

    • in a separate room, the area of ​​\u200b\u200bwhich must be at least 6 m2;
    • in room dental office at the dental chair. In this case, the cabinet area must be at least 14 m2;
    • 2 X-ray machines in one room - the area of ​​the office must be at least 12 m2.

    The room where the X-ray machine will be placed will be called the X-ray room.

  • 7

    Obtaining a ventilation examination protocol. Based on the availability of the necessary SRO permits (Certificate No. 01012.01-2014-5032272285-P-178), we perform this type of work.

  • 8

    Obtaining a sanitary and epidemiological conclusion (SEZ) for the project x-ray room.

    Obtaining a sanitary-epidemiological conclusion for the installation of an X-ray machine is mandatory. On the basis of the Federal Law of 09.01.96 No. 3-FZ "On Radiation Safety of the Population" and the Federal Law of 30.03.99 No. 52-FZ "On the Sanitary and Epidemiological Welfare of the Population" without the permission of the State Sanitary and Epidemiological Service to install X-ray equipment, the office is closed , and its owner is subject to administrative liability in the form of a fine.

  • 9

    Obtaining an act to hide the work. The document provides Building company putting the object into operation.

  • 10

    Registration of documents that are formed in the process of obtaining a SEZ:

    • Technical passport for X-ray room;
    • Instruction on labor protection, including requirements for radiation safety, for the prevention and elimination of radiation accidents;
    • Control and technical journal for the X-ray machine;
    • Order on the assignment of employees to the personnel of groups "A" and "B";
    • Order on the appointment of persons responsible for radiation safety, accounting and storage of X-ray machines, industrial radiation control;
    • Document on personnel training in radiation safety;
    • The conclusion of the medical commission on the passage of personnel of group "A" preliminary and periodic medical examinations;
    • Journal of registration of briefing at the workplace;
    • Accounting cards for individual exposure doses of personnel;
    • Documents confirming the accounting of individual radiation doses of patients.

    The X-ray room is taken into operation by a commission consisting of representatives of the medical institution and the X-ray department, specialists of the sanitary and epidemiological service.

    The sanitary-epidemiological conclusion is a permission for the right to operate an X-ray room.

SanPin describes Medical and Dental X-Ray Units

X-ray equipment

Price, rubles

1. X-ray fluorography apparatus with a fluorescent screen and optical image transfer, film and digital

2. X-ray fluorography low-dose apparatus with a scanning line of detectors and digital image processing

3. X-ray fluorography low-dose device with URI, CCD-matrix and digital image processing

4. X-ray diagnostic apparatus with digital information processing

5. X-ray diagnostic complex with a full set of stands (1st, 2nd and 3rd workstations)

6. X-ray apparatus for fluoroscopy (1st workplace - turntable-support PSSh)

7. X-ray machine for radiography (2nd and 3rd workstations - table of pictures and rack of pictures)

8. Angiographic complex

9. X-ray computed tomography

10. Surgical mobile device with URI

11. Ward X-ray machine

12. X-ray table

13. X-ray machine for lithotripsy

on request

14. Mammography X-ray machine

15. X-ray machine for radiotherapy planning (simulator)

on request

16. Apparatus for close-range X-ray therapy

on request

17. Apparatus for long-range radiotherapy

on request

18. Whole body osteodensitometer

on request

19. Bone densitometer for extremities

on request

20. Osteodesitometer for the whole body and its parts using a wide beam of radiation and a two-dimensional digital detector

on request

Dental units

It is not allowed to use X-ray machines and carry out work not specified in the sanitary and epidemiological conclusion.

Estimated cost of services for putting an X-ray room into operation (development of a placement project, licensing activities in the field of using IRS)

Name of service

Price

Standard (basic) project for an x-ray room with one x-ray machine

from 30 000 rub.

Project for an x-ray room with 2 or more x-ray machines in one room

+50% of the base price for each x-ray machine

Coordination of the project in the authorized body* or Expert opinion on the project for placement of inspection equipment**

from 10 000 rub.

Construction and installation works for the Republic of Belarus (on the basis of an agreed project)

individually

Technical report of the electrical laboratory (contains the protocols of all tests and measurements of the electrical network and equipment)

from 15 000 rub.

The act of inspection of ventilation with the calculation of the level of air exchange in the room

from 15 000 rub.

Protocols for taking parameters by the radiation monitoring laboratory

from 20 000 rub.

Expert opinion on activities in the field of the use of IRS (EZ).

from 60 000 rub.

Sanitary and epidemiological conclusion (SEZ) for medical activities

(additional complex service)

from 80 000 rub.

License for medical activity

(additional complex service)

from 80 000 rub.

State fee for granting a license for medical activities>

(additional complex service)

7 500 rub.

Additionally, specialized services provided by our CA:

Advanced training of your specialists in the programs: "Radiation Safety", "Radiation Control" and others

from 7 000 rub. / person

* - authorized body: Moscow - GORRO, Moscow region - MONIKI, Regions - the relevant regional X-ray and radiological departments (RRO) or other organizations similar in function to the RRO of the region.

** - inspection body - FBUZ "Center for Hygiene and Epidemiology" on a territorial basis (at the location of the X-ray room.

The cost of design, testing and measurement services is calculated individually based on a preliminary assessment of the room configuration, and depends on the number of X-ray equipment to be placed, as well as the type, power and security of this equipment.

The cost of services for obtaining licenses, certificates and conclusions is calculated individually, based on the availability of a basic package of documents, the desired time frame and the complexity of the tasks. .

The X-ray room should have:

  • Stationary means of protection for the procedural and control room, providing attenuation of ionizing radiation (walls, floor, ceiling, protective doors, viewing windows, screens, etc.);
  • Protective measures to prevent scattered radiation from entering other rooms;
  • Bactericidal installations for air disinfection;
  • Closed decorative panels for radiators, providing free air convection and the convenience of wet cleaning;
  • Autonomous supply and exhaust ventilation system;
  • Light board (signal) "Do not enter!" white-red color at the entrance to the treatment room at a height of 1.6-1.8 m from the floor or above the door, automatically lighting up.

According to SanPiN 2.6.1.1192-03 (Appendix No. 7), in order to obtain an Expert Opinion (EP) and a Sanitary and Epidemiological Conclusion (SEZ) for a project and for the implementation of X-ray diagnostic activities, the Technological Design for an X-ray room must be approved by the SPC of Meradiology ( city ​​radiological department (GORRO+] and in Rospotrebnadzor.

Sanitary and epidemiological rules and regulations

SanPiN 2.6.1.1192-03 "Hygienic requirements for the design and operation of x-ray rooms, apparatus and x-ray examinations"

Head of the Department of Radiation Diagnostics of the Central Medical Hospital No. 119, doctor of the highest qualification category, work experience in the specialty for more than 30 years.
Schedule:
Monday-8.00-16.30, Tuesday-13.00-19.00, Wednesday, Thursday-8.00-15.00, Friday - 8.00-19.00
Radiologist Rudneva T.V. . , 25 years of professional experience.
Schedule: Monday, Wednesday, Thursday - 14.00 - 19.30,Tuesday - 8.00 - 13.30.

X-ray examinations are performed in the office:
- radiography of the musculoskeletal system,
- X-ray of the chest organs (on film and digital),
- mammography,
- radiography of teeth (sighting),
- orthopantomography,
- bone densitometry (dual-energy absorptiometry of the lumbar spine and proximal femur).

X-RAY MAMMOGRAPHY IS THE LEADING METHOD IN THE EARLY DIAGNOSIS OF BREAST DISEASES.
Breast cancer (BC) is the main oncological cause of death in Russian women.The only way to overcome the high mortality rate from breast cancer is breast screening, the main purpose of which is to diagnose breast cancer in the early stages and identify benign diseases for timely treatment.Every woman over 35 should be referred for an X-ray mammogram.The mammary gland is a hormonally dependent organ. The most favorable period for the study is from 6 to 12 days from the beginning menstrual cycle. Women in menopause can be tested at any time.
The main advantages of mammography:
- the possibility of polypositional examination of the mammary gland,
- the possibility of visualization of non-palpable formations (in the form of a node, local heavy restructuring of the structure, accumulation of microcalcifications).
Disadvantages of mammography:
- decrease in the information content of the method with a dense background of the mammary gland,
- Difficulties in recognizing metastases in regional lymph nodes.

An additional method for examining the mammary glands is ultrasound method.
If pathological changes are detected during mammography, then the woman should be referred to a mammologist or to an oncology dispensary for a clarifying diagnosis.

THE ROLE OF BONE DENSITOMETRY IN THE DIAGNOSIS AND PREVENTION OF OSTEOPOROSIS
The last decades of the 20th century have shown that diseases of the bones and joints occupy an increasing proportion among the diseases of people over 50 years of age. The aging process of the population of developed countries is considered as one of the main reasons for the predicted increase in the prevalence of inflammatory and degenerative-metabolic diseases in the population. The presence of pathology of the osteoarticular apparatus in people significantly worsens their quality of life due to constant pain and violations of functional activity, thereby burdening the life of not only the patient himself, but also his family, society as a whole.
Among the diseases of the musculoskeletal system, a special place is given to osteoporosis (OP) - one of the most common diseases of the human musculoskeletal system in the world, which is characterized by a decrease in bone mass, a violation of its micro-architectonics and leads to a decrease in bone strength, an increase in the risk of fractures of peripheral bones and spine, which cause temporary disability, disability, as well as increased mortality.
Since 2006 in the central polyclinic of TsMSCh No. 119, a study of bone tissue density is carried out using two-energy and two-photon absorptiometry with an X-ray bone densitometer, which allows the diagnosis of osteoporosis or osteopenia. The remaining densitometric techniques are adapted for the study of peripheral bone sections and therefore can only be used for a preliminary diagnosis.
All methods of densitometry are non-invasive and safe for patients, however, they differ significantly in accuracy, speed, cost of research, and diagnostic information content.
Evaluation options:
- assessment of the proximal femur - allows you to assess the risk of fractures at these critical points,
- spine - assessment of bone mineral density in the lumbar region,
- forearm.
An existing fracture doubles the risk of future fractures, regardless of bone mineral density.
Osteoporosis is a disease that is widespread among the population, especially in the elderly, often leads to a tragic end. Therefore, timely diagnosis, methods of prevention and treatment are of decisive importance.

Welcome to the comprehensive training section: design, build / repair, receive permits to the x-ray room. We offer you services for the preparation, development of technological projects for x-ray rooms and laboratories at the lowest cost and within a reasonable timeframe, we organize laboratory tests necessary to obtain permits. The main activity is obtaining an act of sanitary and epidemiological examination and an expert opinion on activities in the field of the use of radiation sources, a sanitary and epidemiological conclusion, a license for generating sources of ionizing radiation with all intermediate and related procedures (preparation of deployment projects, conducting dosimetric studies, acceptance tests electrical networks of premises, measurement of physical factors, etc.). Those. a set of activities from the moment of receipt terms of reference for the design of the premises to the final stage - obtaining a license for handling generating sources.

We invite to cooperation both manufacturers (suppliers) of equipment, and owners of X-ray rooms and laboratories using X-ray machines for their own purposes:

    1. Prepare a design and working documentation for the construction of checkpoints, customs inspection zones, medical, laboratory and research centers, incl. section "Technological solutions" for the placement of X-ray equipment, with the calculation of radiation protection of workplaces and adjacent premises on the basis of the license of Rospotrebnadzor for III No. 77.99.15.002.L.000023.03.18 dated 03.23.2018

The project is required for all stationary and portable installations that are operated in premises, workshops, offices.

Examples of such settings are:

  • X-ray medical devices, incl. dental aiming, orthopantomograph, dental computed tomography, diagnostic, mobile ward, surgical, mammography, fluorography, densitometer, angiograph, computed tomography);
  • x-ray machines for the treatment of animals (veterinary medicine);
  • non-destructive testing apparatus for X-ray flaw detection;
  • devices for x-ray structural and x-ray spectral analysis;
  • devices for X-ray fluorescence analysis;
  • X-ray microscopes, microprobes and microanalyzers;
  • installations technological control(X-ray thickness gauges, density gauges, level gauges, separators);
  • x-ray machines for screening luggage, hand luggage, bulky cargo, postal correspondence;
  • devices for personal screening of people (X-ray human scanners);
  • linear accelerators of charged particles.

X-ray room design: initial required data, including calculation of radiation protection:

  • Information about materials and thicknesses of enclosing structures (walls, floors, ceilings, doors, etc.).
  • Plan of the premises of the BTI with the designation of the desired location of the apparatus.
  • Information about the functional purpose of adjacent rooms to the treatment room vertically and horizontally, as well as adjacent territories, if any.
  • The height of the treatment room, as well as the height of the rooms adjacent to the treatment room vertically.
  • Passport for x-ray machine.
  • Any other information, the need for which arose in the process of performing work under the contract, solely for the purpose of performing design work.

The composition of the initial data can be specified in the process of developing a project for placement on x-ray equipment. If it is impossible to prepare the initial data, the designer can visit the site.

The cost of developing a project with one X-ray machine is from 30,000 rubles.

After the installation of X-ray equipment in order to subsequently obtain an expert opinion (EC) and a sanitary-epidemiological certificate (SEZ), it is necessary to test the elements of the electrical network and examine the air exchange rate for the supply and exhaust ventilation.

  1. We will carry out acceptance tests of the electrical installation of the building in the premises where X-ray machines are located. The cost of a technical report (inspection of the electrical network) of the premises - from 15,000 rubles.
  2. We will check the ventilation efficiency in the rooms where X-ray machines are located, calculate the air exchange rate for the respective premises, based on the results of the surveys, and prepare a Protocol for examining ventilation installations for the sanitary and hygienic effect with the air exchange rate for the premises.

The cost of the ventilation examination protocol (air exchange rate) of the premises is 15,000 rubles.

All of the above works are carried out on the basis of a SRO certificate, a license for IRS, a certificate of registration of an electrical laboratory up to 35 kV.

We will carry out the preparation and commissioning of the X-ray room on a turnkey basis, you only need to decide on the X-ray equipment that will be operated and its location in the room.

The scope of work that we provide in the process of preparing for the opening of the X-ray room:

  • X-ray equipment placement project with the calculation of radiation safety and subsequent approval by the authorized body;
  • Repair and construction work to ensure protection against x-ray radiation according to the agreed project. The work includes both the use of barite plaster to protect walls, floors and ceilings, as well as the installation of special X-ray protective structures and products: X-ray protective doors, X-ray protective screens, X-ray protective fences, windows and shutters.
  • Conducting tests and measurements of the electrical network and equipment with the provision of a technical report by an accredited electrical laboratory.
  • Examination of ventilation with the calculation of the level of air exchange with the provision of an appropriate protocol.
  • Registration of a package of documents for obtaining an Expert Opinion (ES) with the subsequent receipt of a Sanitary and Hygienic Conclusion (SES).

Designing, building and opening an x-ray room, as well as creating x-ray protection will be an easy task for you in cooperation with our company.

The placement of the X-ray room and equipment in it is carried out on the basis of a project carried out by an organization that has a license for the placement of sources of ionizing radiation (generating) and the design of radiation protection equipment.

The X-ray room, unlike other rooms of medical institutions, is not one room, but several rooms (rooms isolated from each other), each of which has its own purpose and corresponds in area to the established regulatory requirements and also meets all other sanitary and hygienic requirements. We will focus on the latter in detail.

The X-ray room includes at least 3 isolated rooms, one of which serves as a treatment room (a room in which work is directly carried out using ionizing radiation sources), the second is a control room (a room in which the equipment that controls the IRS is located) and the third is an office doctor. Depending on the purpose of the x-ray equipment or the specifics of the activities of the medical institution, the x-ray room may include a photo lab, waiting room, dressing room, staff room, etc. (detailed information about the types of X-ray rooms is given in the article "").

Appendix 5 to SanPiN 2.6.1.1192-03 establishes the composition and area of ​​the X-ray rooms. So, based on the information in Appendix 5 to SanPiN 2.6.1.1192-03, we give below in table 1 the area of ​​​​the procedure room with different x-ray machines, in table 2 the composition and area of ​​​​the premises of the x-ray diagnostic room, in table 3 - the composition and area of ​​\u200b\u200bthe rooms of the radiotherapy room.

Table 1

Treatment room area with different x-ray machines

X-ray machine Area, sq. m (at least)
The use of a wheelchair is envisaged Wheelchair use not included
X-ray diagnostic complex (RDC) with a full set of racks (PSSH, imaging table, imaging rack, imaging rack) 45 40
RDK with PSSH, rack of shots, tripod of shots 34 26
RDK with PSSh and universal stand-tripod, X-ray diagnostic apparatus with digital image processing 34 26
RDK with PSH having remote control 24 16
Apparatus for X-ray diagnostics by X-ray method (image table, image stand, image stand) 16 16
Apparatus for X-ray diagnostics with a universal stand-tripod 24 14
Apparatus for close-range X-ray therapy 24 16
Apparatus for long-distance radiotherapy 24 20
Mammography machine 6
Apparatus for osteodensitometry 8

table 2

The composition and area of ​​​​the premises of the X-ray diagnostic room

The name of a room Area, sq. m
(at least)
General premises of the department (office)
Department manager's office 12
staff room 10 (+3.5 sq. m. for each additional employee)
Results Review Room (Snapshots) 6
Barium Cooking Cabin 3
expected 6
Material 8
Spare parts pantry 6
Storage room for cleaning items 3
Temporary storage room for x-ray film (no more than 100 kg) 6
Staff personal hygiene room 3
Restrooms for staff and patients 3 per cab
Computer 12
Engineering 12
X-ray room
Fluorography room for mass examinations

– procedural
- dressing room
- expected
— photo lab**
- staff room

14
6
6
6
9

Fluorography room for diagnostic images

– procedural
- control room (in the absence of a protective cabin)
— photo lab**
– dressing room*
— doctor's office (for devices with digital image processing)

14
6
6
3
9

X-ray diagnostic room using fluoroscopy and radiography (1,2 and 3 r.m.)

– procedural 1
– procedural 2
- control room
– dressing room*
— photo lab**
- doctor's office

According to the table one
according to the table one
6
3
8
9

Room for X-ray diagnostics of diseases of the gastrointestinal tract intestinal tract(1 r.m.)

– procedural
- control room
— photo lab
- toilet for patients

- doctor's office

According to the table one
6
8
3
4
9

Room for X-ray diagnostics using X-ray and / or tomography (1, 2 and 3 r.m.)

– procedural 1
- control room
– dressing room*
— photo lab**
- staff room

According to the table one
6
3
8
9

Room for X-ray diagnostics of breast diseases using mammography

– procedural
- procedural spec. methods (if necessary)
– dressing room*
— photo lab**
- doctor's office

6
8
3
8
9

Room for X-ray diagnostics of diseases of the genitourinary system (urological)

– treatment room with drain
- control room
— photo lab**
– dressing room with a daybed*
- doctor's office

According to the table one
6
8
4
9

Cabinet (box) for X-ray diagnostics of infectious departments

- vestibule at the entrance to the box (gateway at the entrance to the box)
- expected
- restroom at the expected
– procedural
- control room
— photo lab**
- doctor's office

1,5
6
3
according to the table one
6
8
9

Topometry room (radiotherapy planning)

– procedural
- control room
– barium preparation cabin
— photo lab**
- doctor's office
- restroom

According to the table one
6
3
8
9
3

X-ray operating unit
1. Block for diagnosing diseases of the heart and blood vessels

– X-ray operating room
- control room
— preoperative
– sterilization*
– room for temporary stay of the patient after the study*
— photo lab**
- doctor's office

48
8
6
8
8
8
9

2. Block for the diagnosis of diseases of the lungs and mediastinum

– X-ray operating room
- control room
— preoperative
– sterilization*
– cytological diagnostics*
— photo lab**
— picture viewing room*
- doctor's office
- nurses' room*
— staff personal hygiene room*
– storage room for dirty linen*

32
8
6
6
6
8
6
9
13
4
4

3. Block for diagnosing diseases of the urogenital system

– X-ray operating room
- control room
— photo lab**
- doctor's office
– Contrast preparation room*
- toilet for patients

26
6
8
9
5
3

4. Block for diagnosing diseases of the reproductive organs (mammary gland)

– X-ray operating room
- control room
— photo lab**
- doctor's office

8
4
6
9

X-ray computed tomography room
1. CT room for head examination

– procedural
- control room
— generator/computer
— photo lab**
- doctor's office

18
7
8
8
9

2. CT room for routine examination

– procedural
- control room
— generator/computer
— photo lab**
- doctor's office
- dressing room
- viewing

22
8
8
8
9
4
6

3. CT room for X-ray surgery

– procedural
— preoperative
- control room
— generator/computer
— photo lab**
- doctor's office
- viewing
- room for the preparation of contrast agents
- toilet for patients
- nurse's room
- engineer's room

36
7
10
8
8
9
10
5
3
12
12

* Not necessary.
** Not needed when using digital radiography and fluorography devices.

Table 3

The composition and area of ​​​​the premises of the radiotherapy room

The name of a room Area, sq. m (at least)
1. Near-distance radiotherapy room
– treatment room with 2-3 emitters 16
– treatment room with 1 emitter 12
- control room 9
- doctor's office (examination) 10
- expected 6
2. Room for long-range radiotherapy
– procedural 20
- control room 9
- doctor's office (examination) 10
- expected 6

The use of premises of a smaller area or a reduced set of premises is possible in cases where the equipment used, the organization of work, the number of personnel, etc. ensure compliance with general hygiene requirements (microclimate, bacterial contamination, sanitary and epidemiological regime, etc.).

According to the hygienic requirements of SanPiN 2.6.1.1192-03, it is advisable to place X-ray rooms centrally, as part of the X-ray department, at the junction of the hospital and the clinic. Separately, X-ray rooms of infectious diseases, tuberculosis and obstetric departments of hospitals and, if necessary, fluorography rooms of emergency departments and outpatient departments can be placed.


As a general rule, the X-ray department (office) is not allowed to be placed in residential buildings and children's institutions. The exception is dental X-ray rooms (devices), the possibility of placing them in residential buildings is regulated by a special chapter of the rules SanPiN 2.6.1.1192-03, as well as the "Basic sanitary regulations ensuring radiation safety (OSPORB-99/2010" SP 2.6.1.2612-10 and Amendment 1 to SP 2.6.1.2612-10. It is not allowed to place dental X-ray machines with a film image sensor in residential buildings.

In dental and X-ray dental offices adjacent to residential apartments, it is allowed to place X-ray diagnostic devices with digital image processing, the total rated workload of which does not exceed 40 mA-min / week. (detailed information on this issue is set out in the article "Placement of X-ray equipment in a dental clinic").

It is allowed to operate X-ray rooms in polyclinics built into residential buildings, if the vertically and horizontally adjacent premises are not residential. It is also allowed to place X-ray rooms in an extension to residential building, as well as in the basement floors, if the entrance to the X-ray department (office) is separate from the entrance to the residential building.

As stipulated in clause 3.3 of SanPiN 2.6.1.1192-03, an X-ray department serving only a hospital or only a clinic should be located in the end parts of the building. In this case, the branch should not be through. Entrances to the X-ray department for inpatients and outpatient departments are separate.

Important: by virtue of clause 3.4 of SanPiN 2.6.1.1192-03, it is not allowed:

  • Place X-ray rooms under rooms where water can flow through the ceiling (pools, showers, latrines, etc.);
  • Place the procedural X-ray room adjacent to the wards for pregnant women and children.

According to paragraph 2 of Appendix 7 to SanPiN 2.6.1.1192-03, the choice of premises that make up the X-ray room (department) is carried out by the administration of the medical institution together with the X-ray department (RRO) (or another organization similar in function to the RRO) of the region and is consistent with institution of sanitary and epidemiological supervision. At the same time, in relation to separate rooms X-ray room by a medical organization must comply with the special requirements established by SanPiN 2.6.1.1192-03. However, in practice, there is no longer such approval of the selected premises with Rospotrebnadzor, and applications for approval of the location of the X-ray room are no longer submitted.


Treatment room - a specially equipped room in the X-ray room, in which an X-ray emitter is located and X-ray examinations or X-ray therapy are carried out.


According to paragraph 3.11 of SanPiN 2.6.1.1192-03, the height of the X-ray treatment room should ensure the functioning of technical equipment, for example, a ceiling mount for an X-ray emitter, a tripod, a television monitor, a shadowless lamp, etc. requires a room height of at least 3 m (2.6 meters according to the newer SanPiN 2.1.3.2630-10). The height of the X-ray treatment room in the case of rotational irradiation should be at least 3 m. ×1.8 m.

Requirements for the area of ​​the procedure room, including depending on the X-ray machine, are given in tables 1, 2 and 3 above. At the same time, as stipulated in paragraph 3.9, the area of ​​​​the procedure room can be adjusted in agreement with the body exercising federal sanitary and epidemiological supervision, taking into account the following requirements:

  • The distance from the workplace of the personnel behind a small protective screen to the walls of the room is at least 1.5 m;
  • The distance from the workplace of the personnel behind a large protective screen to the walls of the room is at least 0.6 m;
  • The distance from the swivel table-tripod or from the image table to the walls of the room is at least 1.0 m;
  • The distance from the photo rack to the nearest wall is at least 0.1 m;
  • The distance from the X-ray tube to the viewing window is at least 2 m (for mammography and dental devices - at least 1 m);
  • Technological passage for personnel between the elements of stationary equipment - at least 0.8 m;
  • The area for placing a gurney for a patient is at least 1.5 × 2 m;
  • Additional area in case of technological need to bring a wheelchair into the treatment room - 6 m 2.

In accordance with clause 3.30 of SanPiN 2.6.1.1192-03, in the procedural room, in addition to the procedural room for fluorography and X-ray surgery, it is planned to install a sink with a supply of cold and hot water. In the treatment room, designed for urological examinations, a viduar should be installed.

In the procedural room for the study of children, the presence of toys (which are washed in a soap and soda solution and disinfected) and distracting design is allowed.

Important: It is not allowed to place equipment in the procedural room that is not included in the project, as well as to carry out work that is not related to X-ray examinations.

Subscribe to us

Send your good work in the knowledge base is simple. Use the form below

Students, graduate students, young scientists who use the knowledge base in their studies and work will be very grateful to you.

Posted on http:// www. all best. en/

Brief description of previous work activity

I, Liliya Yagfarovna Musina, born in 1980, graduated from Nizhnekamsk Medical College in 2006 with a degree in Nursing.

In 2001, I got a job at the Municipal Unitary Enterprise "NGMB No. 3" as a medical registrar at the Mammological Center.

At the end of the NMC, she was transferred to the position of a nurse in the same department.

From January 22, 2008 to April 22, 2008, she specialized in KMC in the cycle "Laboratory work in radiology", was transferred to the position of X-ray laboratory assistant and to polyclinic No. 2. In 2009 temporarily worked (for the period of study) in the fluorography room in the prevention department.

From October 1, 2009, to the present, I have been working as an X-ray laboratory assistant in the X-ray room of the Radiation Diagnostics Department of the Consultative and Diagnostic Polyclinic of the State Autonomous Institution "NTsRMB".

General medical experience - 12 years.

Special experience - 5 years

Brief description of the medical institution

In 2009, the health care of Nizhnekamsk has undergone major changes. Since October 1, all adult medical institutions in Nizhnekamsk have merged into a new entity- GAUZ "Nizhnekamsk Central District Multidisciplinary Hospital". The rural part of the Central District Hospital was connected with the district hospital of the river. n. Kamskiye Polyany.

The population of the city is served by two official medical institutions: the Children's City Hospital with a perinatal center (combining the Women's Consultation, the maternity hospital, the children's city hospital) and the Nizhnekamsk Central District Multidisciplinary Hospital, which now includes the former MBUZ "Dental Polyclinic No. 1 and No. 2", MBUZ "City Hospital No. 2", MBUZ "Nizhnekamsk Ambulance Station", the city part of the Central District Hospital, MUE "NGMB No. 3".

The structure of the GAUZ "Nizhnekamsk Central District Multidisciplinary Hospital":

1) The outpatient part includes:

Polyclinic No. 1 (attached population 101467 people)

Polyclinic No. 2 (attached population 86246 people)

Dental clinic (350 visits per shift)

· Consultative and diagnostic polyclinic (500 visits per shift).

In polyclinics No. 1 and No. 2, general practitioners conduct appointments.

Doctors of narrow specialties conduct appointments at the KDP. The Department of Radiation Diagnostics is part of the Consultative and Diagnostic Polyclinic, has firmly taken an authoritative place in the hospital and plays a very important role. important role in the treatment-diagnostic and preventive process.

The structure of the consultative and diagnostic polyclinic includes:

department of narrow specialists No. 1 (therapeutic profile),

department of narrow specialists No. 2 (surgical profile),

department of narrow specialists No. 3

department of radiation diagnostics

Department of functional diagnostics (FGDS, ultrasound, neurophysiology and other functional studies),

clinical diagnostic laboratory,

bacteriological laboratory,

department of medical examinations,

Uro-andrology department

Medical genetic consultation

breast center,

department of prevention

trauma center

2) The stationary part includes 30 divisions. The hospital consists of several buildings, interconnected by elevated passages, which allow transporting patients.

3) The administrative and economic part includes 20 divisions.

The regular number of beds in the hospital is 1110.

Today, 2572 employees work in the hospital. Including: doctors - 432; paramedical workers - 1344; junior medical staff - 480; other personnel - 316.

Department of Radiation Diagnostics

The quality of medical care for the population is determined not only by the adequacy of the forms of its organization, the state of the material and technical base of the healthcare institution, but, first of all, by the availability of qualified specialists. This can have a positive significance only if their work is well organized, materially equipped, and they are economically interested in the results of their work.

Table No. 1. States of health workers in the department of radiation diagnostics

no./p.

Position

By state

Head of department

Senior X-ray Technician

Medical Registrar

Sister is the owner

X-ray room

Radiologist

X-ray laboratory assistant

Nurse

X-ray computed tomography room

Radiologist

X-ray laboratory assistant

Nurse

Cabinet of nuclear magnetic resonance imaging

Radiologist

X-ray laboratory assistant

Nurse

TOTAL for the Department of Radiation Diagnostics:

w.h. medical staff:

average staff

junior staff

other personnel

Table No. 2. List of X-ray diagnostic devices of the Department of Radiation Diagnostics

p/p

Name of equipment

Type, brand

Factory number

Year of issue

X-ray machine with a remotely controlled stand-table "Baccara"

X-ray apparatus with digital radiological

FLASH 2000 system with subtraction

Device for fluorography

Renex-Fluoro

Mobile x-ray machine

"Mobi-Ren-MT"

X-ray mammograph

Computerized X-ray mammograph

Mammo-4-MT

X-ray diagnostic complex. stationary

X-ray complex

diagnostic teleoperated

"KRT-electron"

X-ray CT scanner

Aguilion 16,

X-ray diagnostic complex

"KRD 50/7-Reneks"

X-ray unit

diagnostic mobile

Mobile x-ray diagnostic machine

Mediroll

X-ray diagnostic portable

Film fluorograph

Film fluorograph

Wall-mounted dental x-ray machine

X-ray machine dental apparatus

X-ray machine dental orthopantomograph

"Odontorama RS 100"

"Electron"

X-ray fluorographic digital apparatus

Renex-Fluoro

X-ray diagnostic complex

Mammography x-ray machine with stereotaxy

Mammo-MT

Tomograph medical magnetic resonance

"Image-1"

Mobile ward X-ray machines Medirol and Mobi Ren-MT are located in the departments of anesthesiology and resuscitation No. 1 and No. 2. In the trauma department there is an X-ray mobile device "10 L-6". These mobile devices allow seriously ill patients to be X-rayed directly on the spot.

In the operating unit there is a mobile OKO RTS-612 unit (“C-arm”) used for R-examination in any projections without moving the patient. Designed for clinical use in vascular surgery, endoscopy, urology, traumatology, orthopedics, general surgery.

In its activities, the department of radiation diagnostics is guided by the regulatory legal acts of the Ministry of Health and Social Development Russian Federation, the Republic of Tatarstan, the requirements of the Federal Law "On Radiation Safety Standards" (NRB-99) and "Basic Sanitary Rules for Ensuring Radiation Safety" (OSPORB-2000).

The list of basic documentation that is available in the department:

Sanitary rules and regulations SanPiN 2.6.1.1192-03.

Basic sanitary rules ensuring radiation safety (OSPORB-2000).

Sanitary rules SP 2.6.1.799-99.

Hygienic requirements for the design and operation of X-ray rooms, apparatus and X-ray examinations

Warning scheme for various types of disasters

Intra-hospital orders

· Journal of familiarization with orders

・Safety Log

Folders with instructions on labor protection

Folders with safety instructions

· Folder with exposure dose cards for group A personnel

Folder with job descriptions

· Folders with technical documentation for operating X-ray diagnostic rooms and X-ray diagnostic equipment.

· Journal of maintenance of medical equipment.

Goals and objectives of the radiology department

The main goal of the radiology department is:

Providing the population of the city of Nizhnekamsk and the Nizhnekamsk region with emergency and planned qualified, the most affordable and guaranteed X-ray diagnostic care for patients in a hospital and polyclinic;

Provision of timely preventive fluorographic and mammographic assistance to the population;

Carrying out medical examination of working citizens and timely detection of probable diseases of the chest and mammary glands in the early stages, to prevent an increase in the number of complications and disability;

Conducting constant monitoring of radiation exposure of patients.

Interaction with other medical and preventiveinstitutions and medical schools

The Department of Radiation Diagnostics interacts with polyclinics No. 1, No. 2, the Department of Laboratory Diagnostics, Functional Diagnostics, departments of the hospital, the epidemiological service, and the territorial center for sanitary and epidemiological surveillance. Provides consultative assistance to doctors of other departments of the hospital on the issues of X-ray diagnostics of patients with various diseases. MUP "NTsRMB" cooperates with other hospitals of our republic, transfers information about identified patients with suspected oncopathology to the cancer center in Kazan. After a complete examination, patients with an unclear diagnosis or if it is difficult to choose a treatment are sent to the republican clinical hospital for further in-depth examination and treatment.

Radiologists and X-ray laboratory assistants of the department undergo specialization and advanced training at the Kazan State medical academy and Kazan Medical College.

General organizational measures

All rooms of the department of radiation diagnostics correspond to " hygiene requirements to the arrangement and operation of X-ray rooms, devices and X-ray examinations”. (SanPin 2.1.6.1192-03). All offices have technical data sheets and sanitary and epidemiological conclusions for the operation of X-ray machines. There are orders: on the assignment of working persons to the personnel of group "A"; on the admission of personnel to work with sources of ionizing radiation (IRS); on the appointment of a person responsible for radiation safety. For all rooms there are acts of checking the grounding of X-ray machines and ventilation of rooms. All X-ray machines are equipped with X-ray clinical dosimeters DRK-1 to measure the radiation doses of patients during research. The measured radiation doses of patients are recorded in the patient register, in the medical history and in outpatient cards.

The department conducts individual quarterly dosimetric monitoring of group A personnel with registration of the results in the "Charts of Accounting for Individual Exposure Doses of Personnel". All offices have departments individual means protection for medical personnel and patients: protective screens, aprons, skirts, aprons, collars, hats, gloves, glasses, lead rubber plates, in accordance with the requirements of SanPiN 2.6.1.1192-03, which are tested for lead equivalent once every 2 years.

The department collects and delivers silver-containing waste: x-ray film and used fixer. After the merger of all adult medical institutions of the city into MUP "NTsRMB", a centralized X-ray archive was created in the Department of Radiation Diagnostics for x-rays in a separate isolated room, taking into account all the requirements of SanPiN and fire safety. X-rays are archived for 5 years

The X-ray room where I work is located in Polyclinic No. 1 on the ground floor and is equipped with a RADREX device. X-ray diagnostic complex, designed to work on a table with a grating with a floating upper section FBT-10A and radiography on a stand with the patient in a standing position. The table of the FBT-10A series is intended for performance of all types of radiographic researches. For ease of patient positioning, it is facilitated by a floating top section mechanism. The horizontal movement of the table top is controlled by a foot pedal.

In combination with the X-ray table, a floor-to-ceiling X-ray tube stand is used. Movement or control of all its components is carried out manually, all components are fixed by means of electromagnetic brakes. Click-stops in end position for moving the X-ray tube closer/further and turning the X-ray tube in the vertical direction facilitates positioning for tabletop exposures, for exposures with grating/rack, for exposures with grids and ensures reproducible positioning.

Movement or control of each section can be done using a control box with control handles.

Also in the X-ray room there is a mammography device MAMMOGRAPH. With the help of this device, we perform mammography of the mammary glands in women over 40 years old. It is an X-ray system with exposure control by protective devices, with high frequency generator, with x-ray tube dual focus, detachable carbon fiber mesh grille, efficient magnification device.

The X-ray room consists of the following rooms:

Treatment area 46.3 m 2 ;

Console 15 m 2;

Photo lab 12.5 m 2;

Ordinatorskaya 18 m 2;

Nursing 18 m 2;

The walls, floors and ceilings of each treatment room are covered with a protective layer of barite (plaster), the doors are made of leaded rubber. There are protective shutters on the windows of the treatment room. A viewing window with leaded glass was installed in the wall between the X-ray treatment room and the control room for observing the patient during X-rays. Near front door in the treatment room, at a height of 1.6 meters from the floor, a light board was installed: “Do not enter”, automatically turns on during R-graphy. The walls of the rooms in the R-office are covered with light-colored oil paint. The floors are covered with linoleum. In the photo lab, the floor is covered with ordinary tiles, and the walls up to 2 meters high are covered with glazed tiles.

The lighting of the R-cabinet is artificial. The photo lab additionally has non-active flashlights with a green light filter.

All cabinets are equipped with: supply and exhaust ventilation providing 3-fold air exchange; central heating. There is sewerage, hot and cold water supply.

Organization of work of the X-ray room

The tasks of X-ray diagnostics are performed by an X-ray team consisting of a radiologist, an X-ray laboratory assistant and a nurse.

In his activities, the laboratory assistant reports to the head of the department, and during work - to the radiologist, and is responsible to him for the technical and economic condition of the R-office.

The X-ray laboratory assistant simultaneously supervises the work of the nurse in the R-office and the middle junior medical staff of the medical departments.

At the workplace, the X-ray laboratory assistant acts as three specialists: an average medical worker, an X-ray technician and a photo laboratory assistant, performing alternately the duties of each of them. The workplace of the X-ray laboratory assistant during the working day is constantly changing.

The working day of the X-ray laboratory assistant begins with planning, taking into account the volume and content of the upcoming studies. The time is specified for participation in medical manipulations during R-scopy and complex special studies, as well as for independent work. An approximate order of patients referred for conventional radiography is drawn up. During the working day, it is implemented by contacting the sisters on duty by telephone. All studies are carried out strictly according to the indications and directions of the attending physicians.

The decision on the need for an X-ray examination is made by the radiologist and puts his signature on the referral.

X-ray laboratory assistant maintains medical accounting and reporting documentation in accordance with approved standards and within the established time limits;

1. Journal of diagnostic work of R-graphy (full name, age, diagnosis, address, place of work, film size and quantity, radiation dose and conclusion - the result of the image description).

2. Diary of R-research (preparation for the annual report).

3. Maintenance log of the R-device.

4. Journal of health education.

5. Safety magazine.

Records of the doses received by patients are recorded in the journal of R-graphy and R-scopy, as well as the medical history with the signature of the radiologist.

A daily record of work on surveys and departments is maintained.

I draw up a monthly report, where, in addition to all this, a record is kept of patients who are on a day hospital, referred from the military registration and enlistment office, and non-resident. From 8 o'clock I accept patients prepared for examination: for survey urography, radiography of the lumbosacral spine, irrigography. From 10:00 to 13:00 the rest of the examinations: images of the chest, skull, PPN, spine, limbs.

Before starting work, I visual inspection x-ray room and x-ray machine. I check the integrity of cables, grounding. I do a trial run and then work according to the plan.

radiation radiological complication patient

Stages of radiography

1. I get acquainted with the patient's medical history, I conduct a brief briefing with the patient. I check the referral with the signature of the radiologist. Above the front door I turn on the "Do Not Enter" sign.

2. I turn on the desired workplace. I set the X-ray mode appropriate for this patient. I insert the cassette into the cassette holder.

3. I put the patient on the table deck, center the central beam on the center of the cassette through the area under study and diaphragm the working beam in strict accordance with the size of the area under study.

4. I protect open areas of the body from unused rays, especially critical organs: gonads, mammary glands, thyroid gland, which are not the object of study.

5. I check the correct stacking and centering of the x-rays on the monitor.

6. I give a command to the patient and turn on the high voltage.

7. I release the patient from protective equipment I pick up the cassette.

8. Photochemical processing of X-ray film through the AGFA processing machine.

9. Marking.

Photochemical processing of exposed X-ray films

Photo processing includes several successive stages:

Preparation of photo solutions;

Manifestation;

Intermediate flush;

fixing;

final flush;

Photo solutions are prepared strictly according to the manufacturer's instructions.

I dilute the developer and fixer in enameled jars. The developer must stand for at least 12 hours. Then I filter and pour into tanks-tanks.

I start processing the film with development. I lower the film, tucked into the frame of of stainless steel, into the developer and hold for 6 minutes, then rinse with water and dip into the fixer. I continue fixing until the milky white coating disappears. Next, I wash clean water. Then I label the x-ray. I stick a strip of paper on the raw picture, where I indicate the date, full name, age, case history and department number. As the depletion of the solutions change. I collect the used fixer solution in canisters. Further, the medical technician delivers silver-containing waste to a centralized collection point in Kazan. At present, I am processing films in a Kodak processing machine.

When performing radiography, I strive to limit the radiation exposure to the examined patients and the staff of the R-room:

1. I control the technical condition of X-ray diagnostic equipment.

2. I constantly improve the methodology and technique of research.

3. I apply the optimal physical and technical conditions of the study, contributing to the reduction of radiation exposure to the subjects without loss of diagnostic information:

a) I carefully block the working beam of R-radiation.

b) I use additional filters to optimize the output exposure dose;

c) try to use optimal modes appropriate for a particular object being examined.

4. I use a highly sensitive X-ray film with intensifying screens.

5. I use protective equipment that limits the area of ​​irradiation of the area under study.

I cover the most vulnerable parts of the body with leaded rubber plates: gonads, mammary glands, thyroid gland, etc.

The x-ray room also has protective equipment for personnel: aprons, skirts, gloves, glasses, collars and headgear made of lead rubber. When conducting R.-study using a portable ward apparatus (in the ARC department), radiation protection is carried out by distance.

All employees are equipped with individual dosimeters, which are monitored quarterly. The control of protective equipment is carried out once every two years by specialists of the NEC "Proton". Carry out annual preventive medical examinations employees to identify contraindications to work with radiation sources.

Documents regulating the work of the X-ray room

Sanitary rules and regulations SanPiN 2.6.1.1192-03.

Hygienic requirements for the design and operation of X-ray rooms, apparatus and X-ray examinations.

Basic Sanitary Rules for Ensuring Radiation Safety (OSPORB-2000).

Sanitary rules SP 2.6.1.799-99.

Journal of Safety Engineering.

Hospital orders.

Alert scheme for various types of disasters.

Instructions for the actions of medical personnel in case of fire.

Number of X-ray examinations in 2012

R-graphy of the chest 2357/2707

R-examinations of bones 1032/3238

Irrigography 86/258

R-graphy of bones and joints 3097/5019

Skull R-graphy 85/159

R-graphy of the paranasal sinuses 1174/1174

Plain urography 14/15

Specialist. styling 23/49

Mammography 1428/5706

Total: 9296/18325

Description of more complex X-ray examinations

1. Excretory urography (excretory) - X-ray examination of the urinary tract after intravenous administration water-soluble contrast agent and its excretion by the kidneys.

This research technique allows you to study the morphological and functional state of the kidneys, ureters and bladder. This study is based on the ability of the kidneys to absorb, concentrate and excrete an intravenous iodine-containing substance.

The study is carried out on an empty stomach after cleaning the body enema. Preliminary test for sensitivity to contrast is performed. At least 40 ml of 60-76% solutions of triiodinated substances are used: urotrast, verografin, omnipaque, urografin and others. Indications - all organic diseases and kidney damage.

Radiography is carried out in the position of the patient on the back using R-film 30x40 cm. The laying is performed in such a way that the kidneys, ureters and bladder are displayed on the radiograph. If the shadow of the bladder is cut off due to the patient's high growth, then it is filmed separately on a film of 18x24 cm. Four radiographs are usually performed during urography. The first picture is taken before the introduction of a contrast agent to review the shadow picture of the abdomen and identify possible radio-positive shadows in it (concrements, calcifications, foreign bodies). After the introduction of a contrast agent, as a rule, three pictures are taken - after 7, 15 and 45 minutes.

In some cases, with sufficient contrasting of the urinary tract on them, the doctor will cancel the 3rd image, or, conversely, prescribe additional urograms later in 1-3 hours if pathological changes are noticed.

X-ray stacking instony bone according to Stanvers

During this examination, the patient lies face down on the table top, the chin is pressed to the chest. The head is turned so that the sagittal plane makes an angle of 45 degrees with the plane of the cassette. Under these conditions, the long axis of the investigated pyramid is located in a plane parallel to the deck of the imaging table. The plane of the physiological horizontal is perpendicular to the cassette. The central beam is directed cranially at an angle of 10 degrees to the center of the cassette. Stenvers stacking X-ray is used to assess the structure of the pyramid of the temporal bone and determine the condition of the internal auditory canal, as well as the inner ear.

Mammography in two projections

The purpose of the image is to study the structure of the mammary gland in cases where any induration is found in it during palpation, as well as during a screening examination of women over 40 years old. During the exposure, the patient is facing the x-ray machine. The height of the cassette holder must be adjusted so that the mammary gland is comfortably located on its surface in the center and the nipple is brought to the contour. The mammary gland should be tightly compressed and we take a picture.

Correctness criteria: nipple on the breast contour, medial or lateral region, skin in the inframammary zone.

Irrigography (colonography).

X-ray examination of the colon after filling its lumen with a contrast medium. The patient comes to the study on an empty stomach with the colon thoroughly cleaned of the contents (for this purpose, cleansing enemas are made the night before and in the morning before the study). To introduce a contrast mass into the large intestine through the anus, the Bobrov apparatus is used. This is a device consisting of a balloon for suspensions of barium sulfate, a rubber tube with an enema tip and a rubber bulb for blowing air into the balloon.

The patient is placed on the table. Insert the tip into the anus behind the sphincter of the rectum. Then the large intestine is gradually filled with a contrasting mass under pressure. When it passes into the descending intestine, the patient turns to the left side. Following the introduced contrast suspension, the volume of which is 400-450 ml, air is injected into the intestine. A suspension of barium sulfate envelops the mucous membrane of the colon, and the air inflates its lumen. The resulting double opacification of the colon creates optimal conditions to detect organic changes in it. After radiography of all parts of the colon, the patient is sent to the toilet for maximum bowel emptying from the injected contrast media. A repeated x-ray examination of a partially collapsed intestine supplements information about its condition.

The described simultaneous double contrasting of the colon allows in the best way to study its position, contours, displacement and straightened mucous membrane over a large area. It creates best opportunities for the detection of neoplasms of the intestine.

The main clinical symptoms of complications and the tactics of medical personnel when they occur are as follows:

In R-studies with intravenous administration of iodine-containing R-contrast agents, due to their individual intolerance during the study, dangerous complications are possible that require urgent medical care. For this purpose, the R-office has a first aid kit, which includes:

Adrenaline 0.1% - 1.0.

Norepinephrine 0.2 - 1.0.

Mezaton 1% - 1.0.

Suprastin 2% - 1.0.

Tavegil 0.1 - 2.0.

Prednisolone 30 mg.

Dexamethasone 4 mg.

Ephedrine 5% - 1.0.

Atropine 0.1% - 1.0.

Eufelin 2?4% - 10.0.

Strofantin 0.025% - 1.0.

Cardiamin 2% - 2.0.

Glucose 40% - 20.0.

NaCl 0.9% - 20.

NaCl 0.9 - 400.

Glucose 5% - 400.

Penicillinase 1,000,000 units

Polyglukin 400.0.

Soda 4% - 200.0.

Air duct.

Kornzang.

Language holder.

Gag.

In cases of syncope, cardiovascular collapse, acute respiratory distress, Quincke's edema, anaphylactic shock, the R-room staff begins to provide medical care, followed by a call to the resuscitation team.

The main clinical symptoms of complications and the tactics of medical personnel when they occur are as follows:

Cardiovascularcollapse - acute vascular insufficiency.

It is caused by a drop in the tone of small vessels, and develops with massive blood loss, trauma, myocardial infarction, poisoning, and anaphylactic shock.

Symptoms: pallor of the skin, anxiety, feeling short of breath, cold clammy sweat, bluish extremities, loss of consciousness, weak thready pulse, rapid breathing, lowering blood pressure, the degree of which characterizes the severity of the collapse.

The patient is laid on a couch with an elevated foot end; loosen buckles and straps, create fresh air.

Call the resuscitation team;

Make injections of cordiamine or caffeine;

2. Acute respiratory distress.

May be due to: laryngospasm, bronchospasm, laryngeal edema, pulmonary edema.

Symptoms: shortness of breath, shortness of breath, agitation, tachycardia, cyanosis, foamy sputum.

Avoid retraction of the patient's tongue;

Call the ARC brigade;

In / injected 5-10 ml of a 2.4% solution of aminophylline per 20 ml of 40% glucose, 60 mg of prednisolone, 20 ml of a 30% solution of sodium thiosulfate, 4 ml of a 1% solution of diphenhydramine;

Breathe pure oxygen

Artificial ventilation of the lungs through a mask.

3. Quincke's edema.

Symptoms: the sudden appearance of extensive hyperemic blisters on the skin of the face and a plaque-like rash throughout the body, a rapidly growing swelling of the face and tongue, a feeling of fear, shortness of breath, runny nose, lacrimation, itching.

0.5 ml of a 0.1% solution of adrenaline per 20 ml of NaCl 0.9%, 2 ml of a 1% solution of diphenhydramine, 90 mg of prednisolone, 20 ml of a 30% solution of sodium thiosulfate, 5 ml of a 2.4% solution of eufillin per 20 ml of 40% glucose solution;

If the condition worsens, call the ARC brigade.

4. Anaphylactic shock is the most severe manifestation of an allergic reaction that occurs with the introduction of drugs, radiopaque substances.

Symptoms: rapidly developing pallor of the skin, a feeling of tightness in the chest, severe weakness, dizziness, headache, shortness of breath, nausea, vomiting, urticaria, Quincke's edema, cyanosis, cold sweat, loss of consciousness, convulsions, involuntary urination and defecation, rapid decrease in blood pressure.

Lay the patient on a couch with a raised leg end, create an influx of fresh air

Immediately call the ARC brigade;

Intramuscularly inject 10% -1.0 caffeine or cordiamine 2.0 ml;

In / inject 120 mg of prednisolone per 20 ml of saline, 0.5 ml of 0.1% adrenaline solution per 20 ml of 0.9% NaCl, 4 ml of 1% solution of diphenhydramine or suprastin 2% -2.0.10 ml 10% calcium chloride solution;

In case of respiratory disorders, inject 1.0 ml of 1% solution of lobeline;

To establish breathing with pure oxygen;

Artificial lung ventilation, closed heart massage according to indications.

When rendering emergency care in anaphylactic shock, speed and accuracy of the appointments are required.

Prevention of these complications:

1. Preliminary testing for patient tolerance of a contrast agent;

2. Patients taking antihistamines (diphenhydramine, pipolfen, suprastin, tavegil) in the evening on the eve of the study and in the morning on the day of the study;

3. Prohibition of drinking liquids in the evening on the eve of the study and eating on the day of the study.

Work on professional development

During her work in the X-ray service, she learned to operate a fluorograph, X-ray machines: Prestige, RUM 20, KRD-50/7-Reneks, KRT Electron, Medirol, RTS-612.

Worked out all the used styling and modes. I pass on my practical skills in radiology to beginner radiologists.

I work closely with radiologists. Discuss current topics on X-ray diagnostics. I keep accounting records. Every year I prepare a report of X-ray examinations, a report of R-examinations by departments. I work systematically to improve my skills. I attend conferences, nursing classes.

In his practical activities I use special literature: "Atlas of styling" (authors: Kishkovsky, Tyunin, Esinovskaya). "Medical X-ray technology" (author: Kishkovsky, Tyunin), "Guide for X-ray laboratory technician" (author: V. Yakovets), "Physico-technical foundations of radiology" (author: Monich).

I conduct health promotion and disease prevention work. I conduct explanatory work among patients about the need for this or that X-ray examination, about radiation protection. I conduct conversations with patients, medical staff about the expediency of a healthy lifestyle.

Conducted lectures and discussions on the following topics:

"The role and importance of fluorography in the early diagnosis of tuberculosis and oncological diseases"

"The need for an X-ray examination"

"Movement is life"

"Smoking is injurious to health"

"Alcoholism is a disease"

"Prevention of AIDS infections"

"Prevention of venereological diseases"

“Influenza Prevention. The Importance of Influenza Vaccination

"Healthy lifestyle".

San. dez. mode in x-ray room e

The X-ray room is disinfected and sanitized in a timely manner. Ventilation for 10 minutes every 1-2 hours, especially after fluoroscopy. Sanitary day - once a month. Treatment of the entire R-cabinet - 2 times a month, with a solution of vinegar (for lead dust). Wet cleaning R-cabinet - 2 times a day.

Processing of the shooting table is carried out by wiping with a 1% solution of chloramine after each patient.

Rag treatment - soaking in 1% chloramine solution for 1 hour, then rinsing under running water, dry, store in a closed container for clean rags.

Treatment of enema tips. Soaking in a 3% solution of chloramine, then immersed in a 0.5% washing solution heated to 50C, washed with a brush, then with running water, dried and sent for sterilization to a centralized sterilization department. Currently, disposable enema tips are used.

After use, the Bobrov apparatus is soaked in a 1% solution of chloramine in an appropriate container for 1 hour, then washed, dried, and stored in a clean container.

All utensils for storage, processing are marked and used strictly for their intended purpose. In addition to chloramine, other disinfectants are also used. means: purzhavel, deochlor, calcium hypochlorite.

In the event of a disaster, fire, there are schemes for notifying employees during working and non-working hours, indicating the responsible persons, their addresses and telephone numbers.

San. dez. we carry out the mode on the basis of these orders:

Order No. 720 dated July 31, 1978 "On improving medical care for purulent surgical diseases and strengthening measures to combat nosocomial infections."

Order No. 408 of July 12, 1989 "On measures to reduce the incidence of viral hepatitis in the country."

Order No. 806 dated October 7, 1992 "On strengthening the organization of measures to combat HIV infection."

Order No. 770 dated 06/10/95 "Sterilization and disinfection of medical devices".

SanPiN 2.1.7.728-99 "Rules for the collection and disposal of waste from healthcare facilities."

M.U. No. 287-113 of the Ministry of Health of the Russian Federation 12/30/98. “On disinfection, pre-sterilization cleaning and sterilization of medical devices”.

findings

During the period of work she mastered the basic and special methods of X-ray examination. I made my own table technical parameters studies of various organs and systems to the apparatus "RADREX", I know photolaboratory processing of exposed films. I know the safety rules (anti-radiation protection of patients and staff, fire and electrical safety). I can take pictures in standard and atypical projections in a gentle mode, avoiding additional trauma to patients during the examination. I have good clinical training, I know how to provide emergency first aid, including the simplest methods of resuscitation.

It is important to know the basics of dosimetry of ionizing radiation and radiation biology, knowledge of the physical foundations, means of protection and features of the interaction of X-rays with body tissues.

Offers.

Improve the material and technical bases of X-ray rooms.

Improve the quality of maintenance of X-ray equipment and carry out repairs in a shorter time.

Achieve a reduction in radiation exposure by: choosing the optimal technical parameters, taking into account the characteristics of the patient, using a highly sensitive film and intensifying screens, preventing marriage during photo processing filmed footage, monitoring the preparation of the patient for the study, etc.

To improve the qualifications of the staff of the X-ray room in order to reduce the duration of examinations and defects in work.

X-ray laboratory assistant: __________________ Musina L.Ya.

Hosted on Allbest.ru

Similar Documents

    X-ray and the history of the discovery of x-rays. Means of individual and collective protection in X-ray diagnostics. Dose loads on the population and personnel during medical X-ray examinations and the main ways of their optimization.

    abstract, added 03/21/2008

    Organization of work, requirements for equipment, equipment and tools of a therapeutic dental office. Examination of the patient, determination of the diagnosis. Rules for filling out the clinical documentation of the outpatient card and the patient's medical history.

    abstract, added 04/28/2011

    Equipment of the X-ray room of the multidisciplinary polyclinic of the Medical Unit "Neftyanik". Methods of diagnostic examinations. Practical skills of a laboratory radiologist, his level of qualification. Administrative and regulatory documents that ensure the work.

    practice report, added 01/28/2014

    Establishment of a norm for the types of effects of ionizing radiation on a person in order to limit it. Radiation safety assurance system for medical X-ray examinations. Classification of categories of exposed persons.

    abstract, added 01/04/2012

    The history of the formation of state statistics in the Russian Federation. Organization of the work of a medical institution. social portrait personnel and features of the work of the information and statistics department. Medical statistics and their definition.

    thesis, added 11/25/2011

    General characteristics of the work of the clinic. Organization, content of the work of the physiological department, equipment of the physiotherapy room. Responsibilities of a Physiotherapist. Description of some methods of treatment; help in emergencies.

    practice report, added 08/03/2015

    Studying the work of a nurse in the treatment room of the inpatient department of the Republican Dermatovenerological Dispensary. Cabinet equipment, disinfection modes and procedures general cleaning. Basic measures for injections.

    practice report, added 07/01/2010

    The role of paramedical personnel in psychotherapy, the tactics of a medical worker in a psychiatric clinic. Psychological aspects of a psychiatric hospital, removal of emotional stress due to illness and mobilization of resources.

    abstract, added 04/10/2012

    Organization of the work of the gynecological department of the MUZ "Central City Hospital". Analysis of the dynamics of the work of nurses in the prevention of post-injection complications for 2008-2010, the level of emotional exhaustion of employees from the characteristics of their work.

    thesis, added 11/25/2011

    Development of algorithms for the actions of a nurse. Adverse reactions and complications in angiographic studies. Characteristics of contrast agents. The structure of the department of radiation diagnostics. Indications and contraindications for CT angiography.